Qualitative Method Verification Report template

A qc sheet in the Method Verification procedure (PR21), mapped to ISO 15189:2022 clause 7.3.2. Fill it in whenever the task occurs. Used by the laboratory team.

Qc sheetISO 15189 · 7.3.2As needed
Preview of the Qualitative Method Verification Report as printed by LabToolkit Pro
Calculated automaticallyRed = outside limits
Your LaboratoryQualitative Method Verification Report | YL-LAB-FRM-21-F02b v1.0

Record: Qualitative Method Verification Report — October 2026

ExaminationGlucose
Method / kitGlucose reagent
Comparator method
Plan ref. (PR21-F01b)REF-26-001

2 × 2 Contingency Table (Enter Counts)

True positives (TP)8
False positives (FP)5
False negatives (FN)6
True negatives (TN)10

Results

Positive samples (TP + FN)14
Negative samples (TN + FP)15
Diagnostic sensitivity %57.1
Diagnostic specificity %66.7
Positive predictive value %61.5
Negative predictive value %62.5
Overall agreement %62.1
Cohen's kappa0.24
Sensitivity claim %91.7
Specificity claim %96.6
Sample numbers adequate?Yes
Sensitivity meets claim?No
Specificity meets claim?No
Discordant results: investigation and resolution
Conclusion (method verified for use / not verified)T. Moyo

Sign-off

RoleNameSignatureDate
Performed by
Reviewed by (Section Head)
Approved for use by (Laboratory Director)

Shaded cells are calculated automatically. Values in red are outside their limits. † marks a corrected value (see Corrections).

Controlled document. Printed copies are uncontrolled unless stamped.Page 1
A4 portrait · first page · showing 0 of 0 rowsExactly as LabToolkit Pro prints it — with your logo and colours

What it records

  • Examination
  • Method / kit
  • Comparator method
  • Plan ref. (PR21-F01b)
  • True positives (TP)
  • False positives (FP)
  • False negatives (FN)
  • True negatives (TN)
  • Positive samples (TP + FN)
  • Negative samples (TN + FP)
  • Diagnostic sensitivity %
  • Diagnostic specificity %
  • Positive predictive value %
  • Negative predictive value %
  • Overall agreement %
  • Cohen's kappa
  • Sensitivity claim %
  • Specificity claim %
  • Sample numbers adequate?
  • Sensitivity meets claim?
  • Specificity meets claim?
  • Conclusion (method verified for use / not verified)

Sign-off

  • Performed by
  • Reviewed by (Section Head)
  • Approved for use by (Laboratory Director)

Related templates