IQC Daily Log template

A log in the Internal Quality Control procedure (PR23), mapped to ISO 15189:2022 clause 7.3.7.2. Fill it in daily. Used by the laboratory team.

LogISO 15189 · 7.3.7.2Daily
Preview of the IQC Daily Log as printed by LabToolkit Pro
Calculated automaticallyRed = outside limits
Your LaboratoryIQC Daily Log | YL-LAB-LOG-23-F04 v1.0

Record: IQC Daily Log — October 2026

Run controls before patient samples. Record actual result and compare with ±2SD and ±3SD limits. Apply Westgard Rules. HOLD patient results if QC fails. Open NC (PR07-F01) for repeated failures.
DayDateTest / AnalyteControl NameLot No.Level (L1/L2/L3)ResultAccepted Range (Mean ±2SD)Westgard Rule ViolatedStatus (PASS/FAIL)Analyst InitialsAction Taken (if failed)NC No. (if opened)Reviewed by
12026-10-01GlucoseT. MoyoL2409-A4.944.99PASSTMNone requiredNC-26-001S. Dlamini
22026-10-02CreatinineR. NaidooL2410-B5.055PASSRNNone requiredNC-26-002A. Mensah
32026-10-03ALTS. DlaminiL2411-C4.915PASSSDNone requiredNC-26-003L. Banda
42026-10-04HaemoglobinA. MensahL2412-D5.075FAILAMInvestigated and repeated: within limitsNC-26-004K. Otieno
52026-10-05PotassiumL. BandaL2413-A5.045.02PASSLBNone requiredNC-26-005T. Moyo
62026-10-06GlucoseK. OtienoL2414-B5.035.01PASSKONone requiredNC-26-006R. Naidoo
72026-10-07CreatinineT. MoyoL2415-C5.085PASSTMNone requiredNC-26-007S. Dlamini
82026-10-08ALTR. NaidooL2416-D4.934.98PASSRNNone requiredNC-26-008A. Mensah

Westgard Rules Quick Reference

RuleActionCriterionInterpretation
1₃ₛREJECTOne result >mean ±3SDExcessive random error
2₂ₛREJECTTwo consecutive results exceed the same ±2SD limitSystematic error
R₄ₛREJECTRange between two levels >4SDRandom error
10ₓREJECT10 consecutive results on same side of meanSystematic error / drift
1₂ₛWARNINGOne result >mean ±2SD (no other rule violated)Normal random variation — continue

Sign-off

RoleNameSignatureDate
Completed by
Reviewed by

Shaded cells are calculated automatically. Values in red are outside their limits. † marks a corrected value (see Corrections).

Controlled document. Printed copies are uncontrolled unless stamped.Page 1
A4 landscape · first page · showing 13 of 13 rowsExactly as LabToolkit Pro prints it — with your logo and colours

What it records

  • Day
  • Date
  • Test / Analyte
  • Control Name
  • Lot No.
  • Level (L1/L2/L3)
  • Result
  • Accepted Range (Mean ±2SD)
  • Westgard Rule Violated
  • Status (PASS/FAIL)
  • Analyst Initials
  • Action Taken (if failed)
  • NC No. (if opened)
  • Reviewed by
  • Rule
  • Action
  • Criterion
  • Interpretation

Instructions on the form

  • Run controls before patient samples. Record actual result and compare with ±2SD and ±3SD limits. Apply Westgard Rules. HOLD patient results if QC fails. Open NC (PR07-F01) for repeated failures.

Sign-off

  • Completed by
  • Reviewed by

Related templates

Clause references are to ISO 15189:2022; check them against your licensed copy of the standard.